Loading ...

P381 – 6″ HV/UHV multitechnique deposition cluster tool MPIE

Application

HV/UHV multitechnique deposition cluster tool for thin film and multilayer deposition at 6″ substrates

Year of delivery

2014

Installation site

MPIE, Düsseldorf, Germany

Design Features

  • HV/UHV multitechnique deposition cluster tool with combination of confocal sputtering, E-Beam evaporation, thermal evaporation and different type of effusion cells.
  • The standard sample size and shape is 6″ wafer.
  • HV deposition chamber with two E-Beam evaporation sources, two 2″ magnetrons, one thermal evaporation source and one low temperature effusion cell installed.
    • Deposition uniformity better than 5% for all depostition sources.
    • All confocal 2″ magnetrons with pneumatic in situ tilting.
    • One E-Beam source as rotary multipocket e-beam with motorized pocket exchange
    • Fully motorized 2 axes sample manipulator with integrated sample shutter and maximal sample temperature well above 600°C.
  • UHV deposition chamber with one E-Beam evaporation source and three high temperature effusion cell installed.
    • Deposition uniformity better than 10% for all depostition sources.
    • One E-Beam source as linear multipocket e-beam with motorized pocket exchange
    • Fully motorized 2 axes sample manipulator with integrated sample shutter and maximal sample temperature well above 750°C.
  • Preparation chamber with ion source option and docking port for sample tranportation container.
  • Load lock chamber with storage for 5 samples, installed lamp heating option and connection port to a laminar flow box.
  • Automatic sample transfer between all chambers via a radial distribution chamber.
  • Integrated bake out system.

Special Features

  • Different sample sizes and shapes can be handled (using different kind of sample holders).

Outer Dimensions

Technical specifications and performance values

General

HV deposition chamber

Size

About 760 mm width, about 800 mm depth, about 860 mm height (D-shape chamber with door)

Material

stainless steel

UHV deposition chamber

Size

About 400 mm diameter, about 950 mm height

Material

stainless steel

Load lock chamber

Size

About 300 mm diameter, about 400 mm height

Material

stainless steel

Preparation chamber

Size

About 200 mm diameter, about 850 mm height

Material

stainless steel

Radial distribution chamber

Size

About 1100 mm diameter, about 450 mm height

Material

stainless steel

Vacuum

HV deposition chamber

Base pressure

< 6 * 10-8 mbar

Pump down time

3 hours to < 10-7 mbar

Chamber pumping

Turbo pumping stage, chamber door differentially pumped by dry foreline pump

Bake out

< 120°C

UHV deposition chamber

Base pressure

< 5 * 10-10 mbar

Pump down time

1.5 hours to < 10-7 mbar

Chamber pumping

Turbo pumping stage with dry foreline pump

Bake out

< 150°C

Load lock chamber

Base pressure

< 10-7 mbar

Pump down time

1 hour to < 10-6 mbar

Chamber pumping

Turbo pumping stage with dry foreline pump

Preparation chamber

Base pressure

< 8 * 10-8 mbar

Pump down time

not specified

Chamber pumping

Turbo pumping stage with dry foreline pump (via radial distribution chamber)

Radial distribution chamber

Base pressure

< 9 * 10-8 mbar

Pump down time

4.5 hours to < 2 * 10-7 mbar

Chamber pumping

Turbo pumping stage with dry foreline pump

Manipulator features

HV deposition chamber

Sample size

diameter max. 6″ substrate

Motion axes

2 motorized axes (manipulator z tranlsation and (continous) rotation of the sample stage)

Pneumatic sample shutter (part of the manipulator head)

Temperatures

Room temperature (not stabilized) up to 650°C at sample

UHV deposition chamber

Sample size

diameter max. 6″ substrate

Motion axes

2 motorized axes (manipulator z tranlsation and (continous) rotation of the sample stage)

Pneumatic sample shutter (part of the manipulator head)

Temperatures

Room temperature (not stabilized) up to 850°C at sample

Sample preparation features

Load lock chamber

Thermal treatment

max. 200°C at sample (no temperature regulation)

Preparation chamber

Oxydation / Nitration /
Plasma treatment

max. 5 * 10-4 mbar partly ionised gas mixture (using a griddless ion gun)
Gas mixture variable from pure argon up to pure oxygen or nitrogen

Ion beam etching /
sample precleaning

Variable ion source to sample distance
Wide range variable ion energy and ion beam current

Performance test results

Chamber pump down (part 1)
Chamber pump down (part 2)