Loading ...
P401 - Plane grating monochromator for soft X-ray synchrotron radiation at Russian-German beamline at BESSY, Helmholtz-Zentrum Berlin, Germany - Bestec GmbH
P401 - Plane grating monochromator for soft X-ray synchrotron radiation at Russian-German beamline at BESSY, Helmholtz-Zentrum Berlin, Germany - Bestec GmbH

P401 – PGM RGBL BESSY

Application

Plane grating monochromator (PGM) for soft X-ray synchrotron radiation

Year of delivery

2015

Installation site

Russian-German beamline at BESSY, Helmholtz-Zentrum Berlin, Germany

Design Features

  • High-resolution grazing incidence plane grating monochromator (PGM).

  • Spectral range: 10 – 1300 eV.
  • Optical elements: one plane mirror and three plane gratings.
  • Temperature sensors at mirror and gratings chin guards.
  • Electrodes at mirror and gratings surfaces for carbon contamination prevention.
  • Provisions for RF oxygen plasma cleaning of the optical surfaces.
  • Active water cooling of all optical elements.

  • Permanently installed bakeout provisions.
  • Adjustable support structure.

Performance Features

  • High positioning resolution (minimum incremental motion).
  • High pitch rotation stability and repeatability.
  • Minimized mounting induced deformations.

Outer Dimensions

P401 - Outer dimensions - Plane grating monochromator for soft X-ray synchrotron radiation at Russian-German beamline at BESSY, Helmholtz-Zentrum Berlin, Germany - Bestec GmbH

Technical specifications and performance values

General

Vacuum chamber

Chamber type

cylindrical, vertical axis

Main flanges

at the top

Material

1.4404 (C < 0.02%)

Base pressure

10-10 mbar range

Optical design parameters

Beam height (entrance)

1400 mm

Beam offset

20 mm

Number of plane mirrors

1

Plane mirror dimensions

310 mm x 30 mm x 60 mm

Number of gratings

2

Gratings dimensions

100 mm x 20 mm x 40 mm

Support structure

Adjustment ranges

Vertical translation

± 10 mm

Lateral translation

± 10 mm

Longitudinal translation

± 10 mm

Pitch

± 1°

Roll

± 1°

Yaw

± 1°

Mirror mechanics

Pitch rotation

Range

-14° to +0.55°

Resolution (design value)

0.004 µrad /motor fine step
(1/64, recommended)
0.00146 µrad /encoder count

Resolution (measured value)

0.50 µrad

Repeatability

0.14 µrad (pitch)

0.25 µrad (roll)

Stability

0.10 µrad

Translation

Range

81 mm

Resolution (design value)

0.05 µm /motor fine step
(1/64, recommended)
0.005 µm /encoder count

Repeatability

0.20 µrad (pitch)

0.30 µrad (roll)

Grating mechanics

Pitch rotation

Range

-19.98° to +1.25°

Resolution (design value)

0.004 µrad /motor fine step
(1/64, recommended)
0.00146 µrad /encoder count

Resolution (measured value)

0.50 µrad

Repeatability

0.07 µrad (pitch)

0.20 µrad (roll)

Stability

0.06 µrad

Translation

Range

300 mm

Resolution(design value)

0.05 µm /motor fine step
(1/64, recommended)
0.005 µm /encoder count

Repeatability

0.20 µrad (pitch)

0.60 µrad (roll)

Performance Test Results

Pitch deviation after pitch change (mirror)
P401 - Pitch deviation after pitch change (mirror) - Plane grating monochromator for soft X-ray synchrotron radiation at Russian-German beamline at BESSY, Helmholtz-Zentrum Berlin, Germany - Bestec GmbH
Roll deviation after pitch change (mirror)
P401 - Roll deviation after pitch change (mirror) - Plane grating monochromator for soft X-ray synchrotron radiation at Russian-German beamline at BESSY, Helmholtz-Zentrum Berlin, Germany - Bestec GmbH
Pitch deviation after pitch change (grating)
P401 - Pitch deviation after pitch change (grating) - Plane grating monochromator for soft X-ray synchrotron radiation at Russian-German beamline at BESSY, Helmholtz-Zentrum Berlin, Germany - Bestec GmbH
Roll deviation after pitch change (grating)
P401 - Roll deviation after pitch change (grating) - Plane grating monochromator for soft X-ray synchrotron radiation at Russian-German beamline at BESSY, Helmholtz-Zentrum Berlin, Germany - Bestec GmbH
Pitch deviation after mirror translation
P401 - Pitch deviation after mirror translation - Plane grating monochromator for soft X-ray synchrotron radiation at Russian-German beamline at BESSY, Helmholtz-Zentrum Berlin, Germany - Bestec GmbH
Roll deviation after mirror translation
P401 - Roll deviation after mirror translation - Plane grating monochromator for soft X-ray synchrotron radiation at Russian-German beamline at BESSY, Helmholtz-Zentrum Berlin, Germany - Bestec GmbH
Pitch deviation after grating translation
P401 - Pitch deviation after grating translation - Plane grating monochromator for soft X-ray synchrotron radiation at Russian-German beamline at BESSY, Helmholtz-Zentrum Berlin, Germany - Bestec GmbH
Roll deviation after grating translation
P401 - Roll deviation after grating translation - Plane grating monochromator for soft X-ray synchrotron radiation at Russian-German beamline at BESSY, Helmholtz-Zentrum Berlin, Germany - Bestec GmbH
Mirror pitch positioning resolution
P401 - Mirror pitch positioning - Plane grating monochromator for soft X-ray synchrotron radiation at Russian-German beamline at BESSY, Helmholtz-Zentrum Berlin, Germany - Bestec GmbH
Grating pitch positioning resolution
P401 - Grating pitch positioning - Plane grating monochromator for soft X-ray synchrotron radiation at Russian-German beamline at BESSY, Helmholtz-Zentrum Berlin, Germany - Bestec GmbH
Residual gas analysis
P401 - Residual gas analysis - Plane grating monochromator for soft X-ray synchrotron radiation at Russian-German beamline at BESSY, Helmholtz-Zentrum Berlin, Germany - Bestec GmbH