Loading ...
P518 - Exit slit unit for soft X-ray synchrotron radiation, ATHOS (Diavolezza) beamline at at Swiss FEL, Paul Scherrer Institut, Switzerland
P518 - Exit slit unit for soft X-ray synchrotron radiation, ATHOS (Diavolezza) beamline at at Swiss FEL, Paul Scherrer Institut, Switzerland

P518 – Exit Slit Unit DIAVOLEZZA SwissFEL PSI

Application

Exit slit unit for soft X-ray radiation

Year of delivery

2025

Installation site

ATHOS beamline at Swiss FEL, Paul Scherrer Institut,  Switzerland

Design Features

  • Vertically defining slit with motorized and encoded slit opening.
  • Symmetric slit opening, non-linear opening characteristics.
  • OFHC slit blades.
  • Motorized and encoded manipulator (horizontal/vertical direction) with holder for scintillator screens.
  • Adjustable support structure.

Performance Features

  • Precision and slit opening repeatability smaller than 150 nm.
  • Parallelism smaller than 0.5 µm over 15 mm edge length.
  • Straightness smaller than 0.2 µm over 15 mm edge length.
  • Very small deviation (submicrometer) of the slit opening calibration curve from the theoretical non-linear slit opening function.

Outer Dimensions

P518 - Outer dimensions - Exit slit unit for soft X-ray synchrotron radiation, ATHOS (Diavolezza) beamline at at Swiss FEL, Paul Scherrer Institut, Switzerland

Technical specifications and performance values

General

Vacuum chamber

Chamber type

cylindrical, horizontal axis

Main flanges

DN160CF cluster flanges

Material

1.4301

Vertically defining slit (VDS)

Slit blades

Material

OFHC

Straightness

< 0.20 µm over 15 mm (1σ)

Parallelism

0.02 μm /mm
(at ca. 50 μm slit opening)
-0.01 μm /mm
(at ca. 100 μm slit opening)
0.03 μm /mm
(at ca. 200 μm slit opening)

Slit opening

Range

4.6 µm to 9.20 mm
(+4 mm at the notch)

Repeatability

< 0.10 µm
(for slit apertures < 300 µm)
< 0.15 µm
(for slit apertures > 300 µm)

Variation of slit center position while opening

< 1 µm

Resolution (design value)

0.00156 µm /motor fine step
(1/64, recommended)
0.05 µm /encoder count

Resolution (measured value)

0.01 µm

Manipulator for scintillator screens

Horizontal translation

Range

-11 mm to 11 mm

Resolution (design value)

0.23 µm /motor fine step
(1/64, recommended)
0.05 µm /encoder count

Resolution (measured value)

0.20 µm

Repeatability

< 0.10 µm (translation)

Vertical translation

Range

-0.15 mm 100.2 mm

Resolution (design value)

0.23 µm /motor fine step
(1/64, recommended)
0.05 µm /encoder count

Resolution (measured value)

0.20 µm

Repeatability

< 0.10 µm (translation)

Performance Test Results

Slit width calibration (VDS)
P518 - Slit width calibration (VDS) - Exit slit unit for soft X-ray synchrotron radiation, ATHOS (Diavolezza) beamline at at Swiss FEL, Paul Scherrer Institut, Switzerland
Variation of slit center position (VDS)
P518 - Variation of slit center position (VDS) - Exit slit unit for soft X-ray synchrotron radiation, ATHOS (Diavolezza) beamline at at Swiss FEL, Paul Scherrer Institut, Switzerland
Slit opening repeatability (VDS)
P518 - Slit opening repeatability (VDS) - Exit slit unit for soft X-ray synchrotron radiation, ATHOS (Diavolezza) beamline at at Swiss FEL, Paul Scherrer Institut, Switzerland
Slit opening repeatability (VDS)
P518 - Slit opening repeatability (VDS) - Exit slit unit for soft X-ray synchrotron radiation, ATHOS (Diavolezza) beamline at at Swiss FEL, Paul Scherrer Institut, Switzerland
Slit blades parallelism (VDS)
P518 - Slit blades parallelism (VDS) - Exit slit unit for soft X-ray synchrotron radiation, ATHOS (Diavolezza) beamline at at Swiss FEL, Paul Scherrer Institut, Switzerland
Slit blades parallelism (VDS)
P518 - Slit blades parallelism (VDS) - Exit slit unit for soft X-ray synchrotron radiation, ATHOS (Diavolezza) beamline at at Swiss FEL, Paul Scherrer Institut, Switzerland
Slit blades parallelism (VDS)
P518 - Slit blades parallelism (VDS) - Exit slit unit for soft X-ray synchrotron radiation, ATHOS (Diavolezza) beamline at at Swiss FEL, Paul Scherrer Institut, Switzerland
Slit blades parallelism (VDS)
P518 - Slit blades parallelism (VDS) - Exit slit unit for soft X-ray synchrotron radiation, ATHOS (Diavolezza) beamline at at Swiss FEL, Paul Scherrer Institut, Switzerland
Slit opening positioning resolution (VDS)
P518 - Slit opening positioning resolution (VDS) - Exit slit unit for soft X-ray synchrotron radiation, ATHOS (Diavolezza) beamline at at Swiss FEL, Paul Scherrer Institut, Switzerland
Slit opening positioning resolution (VDS)
P518 - Slit opening positioning resolution (VDS) - Exit slit unit for soft X-ray synchrotron radiation, ATHOS (Diavolezza) beamline at at Swiss FEL, Paul Scherrer Institut, Switzerland
Slit opening positioning resolution (VDS)
P518 - Slit opening positioning resolution (VDS) - Exit slit unit for soft X-ray synchrotron radiation, ATHOS (Diavolezza) beamline at at Swiss FEL, Paul Scherrer Institut, Switzerland
Slit opening positioning resolution (VDS)
P518 - Slit opening positioning resolution (VDS) - Exit slit unit for soft X-ray synchrotron radiation, ATHOS (Diavolezza) beamline at at Swiss FEL, Paul Scherrer Institut, Switzerland
Slit opening positioning resolution (VDS)
P518 - Slit opening positioning resolution (VDS) - Exit slit unit for soft X-ray synchrotron radiation, ATHOS (Diavolezza) beamline at at Swiss FEL, Paul Scherrer Institut, Switzerland
Slit opening positioning resolution (VDS)
P518 - Slit opening positioning resolution (VDS) - Exit slit unit for soft X-ray synchrotron radiation, ATHOS (Diavolezza) beamline at at Swiss FEL, Paul Scherrer Institut, Switzerland
Manipulator horizontal deviation after translation
P518 - Manipulator horizontal deviation after translation - Exit slit unit for soft X-ray synchrotron radiation, ATHOS (Diavolezza) beamline at at Swiss FEL, Paul Scherrer Institut, Switzerland
Manipulator horizontal positioning resolution
P518 - Manipulator horizontal positioning - Exit slit unit for soft X-ray synchrotron radiation, ATHOS (Diavolezza) beamline at at Swiss FEL, Paul Scherrer Institut, Switzerland
Manipulator vertical deviation after translation
P518 - Manipulator vertical deviation after translation - Exit slit unit for soft X-ray synchrotron radiation, ATHOS (Diavolezza) beamline at at Swiss FEL, Paul Scherrer Institut, Switzerland
Manipulator vertical positioning resolution
P518 - Manipulator vertical positioning - Exit slit unit for soft X-ray synchrotron radiation, ATHOS (Diavolezza) beamline at at Swiss FEL, Paul Scherrer Institut, Switzerland
Residual gas analysis
P518 - Residual gas analysis - Exit slit unit for soft X-ray synchrotron radiation, ATHOS (Diavolezza) beamline at at Swiss FEL, Paul Scherrer Institut, Switzerland